Development of a single endovascular device for aortic valve replacement and ascending aortic repair.
Journal of Cardiac Surgery 2014 May
BACKGROUND AND AIM OF THE STUDY: Transcatheter aortic valve implantation (TAVI) is contraindicated in the presence of an ascending aortic aneurysm. Our aim was to design a composite endovascular device enabling ascending aortic repair and TAVI.
METHODS: From 2007 to 2013, among 1196 patients with severe aortic stenosis screened for TAVI, 79 nonbicuspid patients had ascending aortic diameter >45 mm. Proximal aortic geometry was assessed in those with computed tomography angiography.
RESULTS: All together, 51 patients (35 males, aged 85 ± 8 years; 19 TAVI, 10 open Wheat procedures, 22 managed conservatively) were included. The required annular diameter for implantation of currently available TAVI prostheses was met in 41% (21/51). Novel prosthetic valves appropriate for annular range up to 30 mm would extend device applicability to 78% (40/51). Proximal and distal diameters of the graft-covering portion ranging between 30 and 46 mm would enable 10% graft oversizing in all but six patients. In 88% (45/51) the required minimum 10 mm distance between aortic valve annulus and coronary artery ostia was found. Mean distance between left and right coronary artery ostia and sinotubular junction was 2.6 ± 1.5 and 3.2 ± 1.7 mm, respectively.
CONCLUSIONS: Novel composite endovascular valved grafts may extend the application of transcatheter techniques to patients denied TAVI due to a concomitant ascending aneurysm. The location of coronary arteries in relation to the sinotubular junction must be addressed in designing these composite valve grafts.
METHODS: From 2007 to 2013, among 1196 patients with severe aortic stenosis screened for TAVI, 79 nonbicuspid patients had ascending aortic diameter >45 mm. Proximal aortic geometry was assessed in those with computed tomography angiography.
RESULTS: All together, 51 patients (35 males, aged 85 ± 8 years; 19 TAVI, 10 open Wheat procedures, 22 managed conservatively) were included. The required annular diameter for implantation of currently available TAVI prostheses was met in 41% (21/51). Novel prosthetic valves appropriate for annular range up to 30 mm would extend device applicability to 78% (40/51). Proximal and distal diameters of the graft-covering portion ranging between 30 and 46 mm would enable 10% graft oversizing in all but six patients. In 88% (45/51) the required minimum 10 mm distance between aortic valve annulus and coronary artery ostia was found. Mean distance between left and right coronary artery ostia and sinotubular junction was 2.6 ± 1.5 and 3.2 ± 1.7 mm, respectively.
CONCLUSIONS: Novel composite endovascular valved grafts may extend the application of transcatheter techniques to patients denied TAVI due to a concomitant ascending aneurysm. The location of coronary arteries in relation to the sinotubular junction must be addressed in designing these composite valve grafts.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app