BACKGROUND: Nexfin® (BMEYE, Amsterdam, The Netherlands) is a totally non-invasive blood pressure and cardiac output (CO) monitor based on finger arterial pulse contour analysis.
METHODS: We performed an open observational study in a mix of medical-surgical-burns critically ill patients (N.=45) to validate Nexfin obtained blood pressures (MAPnex) against PiCCO (MAPfem) derived blood pressure measurements. MAPnex, MAPfem and corresponding systolic (SBP) and diastolic (DBP) blood pressures were measured continuously and registered with a 2 hour interval during the 8-hour study period. Statistical analysis was performed by Pearson regression, Bland and Altman, Concordance plot and Polar plot analysis.
RESULTS: MAPnex shows excellent correlation with MAPfem (R² 0.88, mean bias ± LA -2.3±12.4 mmHg, 14.7% error) and may be used interchangeably with invasive monitoring. The excellent MAPnex -MAPfem correlation was preserved in subgroup analysis for patients with severe hypotension, high systemic vascular resistance, low CO, hypothermia and in patients supported by inotropic/vasopressive agents. MAPnex is able to follow changes in MAPfem during the same time interval (level of concordance 85.5%). Nexfin SBP and DBP show significant correlation with PiCCO but the criteria for interchangeability were not met. Finally, polar plot analysis showed that trending capabilities were excellent when changes in MAPnex (ΔMAPnex) were compared to ΔMAPfem (96.1% of changes were within the level of 10% limits of agreement).
CONCLUSION: In this sample of critically ill patients we found a good correlation between MAPnex and invasive blood pressures obtained by PiCCO.
METHODS: We performed an open observational study in a mix of medical-surgical-burns critically ill patients (N.=45) to validate Nexfin obtained blood pressures (MAPnex) against PiCCO (MAPfem) derived blood pressure measurements. MAPnex, MAPfem and corresponding systolic (SBP) and diastolic (DBP) blood pressures were measured continuously and registered with a 2 hour interval during the 8-hour study period. Statistical analysis was performed by Pearson regression, Bland and Altman, Concordance plot and Polar plot analysis.
RESULTS: MAPnex shows excellent correlation with MAPfem (R² 0.88, mean bias ± LA -2.3±12.4 mmHg, 14.7% error) and may be used interchangeably with invasive monitoring. The excellent MAPnex -MAPfem correlation was preserved in subgroup analysis for patients with severe hypotension, high systemic vascular resistance, low CO, hypothermia and in patients supported by inotropic/vasopressive agents. MAPnex is able to follow changes in MAPfem during the same time interval (level of concordance 85.5%). Nexfin SBP and DBP show significant correlation with PiCCO but the criteria for interchangeability were not met. Finally, polar plot analysis showed that trending capabilities were excellent when changes in MAPnex (ΔMAPnex) were compared to ΔMAPfem (96.1% of changes were within the level of 10% limits of agreement).
CONCLUSION: In this sample of critically ill patients we found a good correlation between MAPnex and invasive blood pressures obtained by PiCCO.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app