COMPARATIVE STUDY
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Syringe plunger contamination by hazardous drugs: a comparative study.

AIM: The purpose of this study was to perform a comparative cyclophosphamide contamination level test with Becton Dickinson® syringe plungers with Phaseal® Closed System Transfer Devices and Equashield® syringe plungers under routine oncological compounding conditions.

METHOD: The ChemoGlo™ sampling kit and analysis services were used to test for cyclophosphamide contamination levels on the syringe plungers of Becton Dickinson® Phaseal and Equashield® syringes that underwent cycles of drug transfer in a Forma Class II, 2A Biological Safety Cabinet. Prior to testing, the syringes were divided into three equal groups for the Equashield® and Becton Dickinson® syringes. A 50 mL aliquot of cyclophosphamide was drawn into each syringe and then injected back into the cyclophosphamide vial. This drug transfer procedure was immediately repeated twice for the syringes in group 1, four times for group 2, and eight times for group 3. After the completion of the drug transfers with the Equashield® and Becton Dickinson® Phaseal syringes, the plungers were retracted back to the nominal syringe marking, and a wipe test of the exposed plunger was done using the ChemoGlo™ sampling kit.

RESULTS: Significant contamination levels of 2000 ng and greater were detected on most Becton Dickinson® syringe plungers with Phaseal® Closed System Transfer Devices, whereas all Equashield® syringes remained uncontaminated at undetectable levels.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app