Health worker exposure risk during inhalation sedation with sevoflurane using the (AnaConDa®) anaesthetic conserving device.

INTRODUCTION AND OBJECTIVE: Occupational exposure to sevoflurane should not exceed 2 ppm. During inhalation sedation with sevoflurane using the anaesthetic conserving device (AnaConDa(®)) in the post-anaesthesia care unit, waste gases can be reduced by gas extraction systems or scavenging devices such as CONTRAfluran™. However, the efficacy of these methods has not been clearly established. To determine the safest scenario for healthcare workers during inhalation sedation with sevoflurane in the post-surgical intensive care unit.

MATERIALS AND METHODS: An experimental study on occupational exposure was conducted in a post-cardiothoracic care unit during March-August 2009. The measurements were performed in four post-cardiac surgery sedated adults in post-surgical intensive care unit and four nurses at the bedside, and at four points: scenario A, inhalation sedation without gas extraction system or contrafluran as a reference scenario; scenario B, applying a gas extraction system to the ventilator; scenario C, using contrafluran; and scenario 0, performing intravenous isolation sedation. Sevoflurane concentrations were measured in the nurses' breathing area during patient care, and at 1.5 and 8 m from the ventilator using diffusive passive monitor badges.

RESULTS: All badges corresponding to the nurses' breathing area were below 2 ppm. Levels of sevoflurane detected using prevention systems were lower than that in the control situation. Only one determination over 2 ppm was found, corresponding to the monitor placed nearest the gas outlet of the ventilator in scenario A. Trace concentrations of sevoflurane were found in scenario 0 during intravenous sedation.

CONCLUSIONS: Administration of sevoflurane through the AnaConDa(®) system during inhalation sedation in post-surgical intensive care units is safe for healthcare workers, but gas extraction systems or scavenging systems, such as CONTRAfluran™ should be used to reduce occupational exposure as much as possible.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app