Reduction in Surface Contamination With Cyclophosphamide in 30 US Hospital Pharmacies Following Implementation of a Closed-System Drug Transfer Device.
Hospital Pharmacy 2013 March
PURPOSE: In a follow-up to a previous study, surface contamination with the antineoplastic drug cyclophosphamide was compared in 30 US hospital pharmacies from 2004 to 2010 following preparation with standard drug preparation techniques or the PhaSeal closed system drug transfer device (CSTD).
METHODS: Wipe samples were taken from biological safety cabinet (BSC) surfaces, BSC airfoils (the front leading edge of the BSC), floors in front of BSCs, and countertops in the pharmacy, and they were analyzed for contamination with cyclophosphamide. Contamination was reassessed after a minimum of 6 months following the implementation of the CSTD. Surface contamination (ng/cm(2)) was compared between the 2 techniques and between the previous and current test periods and evaluated with the Kruskal-Wallis test.
RESULTS: With the use of CSTD compared to the standard preparation techniques, a significant reduction in levels of contamination with cyclophosphamide was observed (P < .0001). Median values for surface contamination with cyclophosphamide were reduced by 86% compared to 95% in the previous study.
CONCLUSIONS: The CSTD significantly reduced, but did not totally eliminate, surface contamination with cyclophosphamide. In addition to other protective measures, increased usage of CSTDs should be employed to help protect health care workers from exposure to hazardous drugs.
METHODS: Wipe samples were taken from biological safety cabinet (BSC) surfaces, BSC airfoils (the front leading edge of the BSC), floors in front of BSCs, and countertops in the pharmacy, and they were analyzed for contamination with cyclophosphamide. Contamination was reassessed after a minimum of 6 months following the implementation of the CSTD. Surface contamination (ng/cm(2)) was compared between the 2 techniques and between the previous and current test periods and evaluated with the Kruskal-Wallis test.
RESULTS: With the use of CSTD compared to the standard preparation techniques, a significant reduction in levels of contamination with cyclophosphamide was observed (P < .0001). Median values for surface contamination with cyclophosphamide were reduced by 86% compared to 95% in the previous study.
CONCLUSIONS: The CSTD significantly reduced, but did not totally eliminate, surface contamination with cyclophosphamide. In addition to other protective measures, increased usage of CSTDs should be employed to help protect health care workers from exposure to hazardous drugs.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app