We have located links that may give you full text access.
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Evaluation of bacterial contamination on surgical drapes following use of the Bair Hugger(®) forced air warming system.
This pilot study determined the rate of bacterial contamination on surgical drapes of small animal patients warmed intra-operatively with the Bair Hugger(®) forced air warming system compared to a control method. Surgical drapes of 100 patients undergoing clean surgical procedures were swabbed with aerobic culturettes at the beginning and end of surgery. Samples were cultured on Trypticase soy agar. Contamination of the surgical drapes was identified in 6/98 cases (6.1%). There was no significant difference in the number of contaminated surgical drapes between the Bair Hugger(®) and control groups (P = 0.47).