We have located links that may give you full text access.
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
MR guidance and thermometry of percutaneous laser disc decompression in open MRI: an ex vivo study.
PURPOSE: To assess the feasibility of guidance and thermometry by open 1.0 T magnetic resonance (MR) imaging during percutaneous laser disc decompression (PLDD).
METHODS: A fluoroscopic proton-density-weighted turbo spin echo sequence was used for positioning a laser fiber and a reference thermosensor within the targeted spinal disc. In 30 lumbar discs from human donors, nonspoiled gradient-echo (GRE) sequences with different echo times (TE) were compared to monitor thermal laser effects (Nd:YAG laser, 1,064 nm). Temperature distribution was visualized in real time on the basis of T1-weighted images and the proton resonance frequency (PRF) technique. Image quality, temperature accuracy, and correlation with macroscopic lesion sizes were analyzed. Image quality was confirmed in healthy volunteers.
RESULTS: MR-guided placement of the laser fiber in the center of the targeted disk was precise. Best overall PLDD results-considering image quality (contrast-to-noise ratio), temperature accuracy (R (2) = 0.96), and correlation between the macroscopic and MR lesions (R (2) = 0.63)-were achieved with TE at 7 ms. The same TE value also gave the best image quality with healthy volunteers.
CONCLUSION: Instrument guidance and PRF-based thermometry of PLDD in the lumbar spine are feasible and accurate. Open 1.0 T MR imaging with fast spin-echo and GRE sequence designs may render laser discectomies more effective and controllable.
METHODS: A fluoroscopic proton-density-weighted turbo spin echo sequence was used for positioning a laser fiber and a reference thermosensor within the targeted spinal disc. In 30 lumbar discs from human donors, nonspoiled gradient-echo (GRE) sequences with different echo times (TE) were compared to monitor thermal laser effects (Nd:YAG laser, 1,064 nm). Temperature distribution was visualized in real time on the basis of T1-weighted images and the proton resonance frequency (PRF) technique. Image quality, temperature accuracy, and correlation with macroscopic lesion sizes were analyzed. Image quality was confirmed in healthy volunteers.
RESULTS: MR-guided placement of the laser fiber in the center of the targeted disk was precise. Best overall PLDD results-considering image quality (contrast-to-noise ratio), temperature accuracy (R (2) = 0.96), and correlation between the macroscopic and MR lesions (R (2) = 0.63)-were achieved with TE at 7 ms. The same TE value also gave the best image quality with healthy volunteers.
CONCLUSION: Instrument guidance and PRF-based thermometry of PLDD in the lumbar spine are feasible and accurate. Open 1.0 T MR imaging with fast spin-echo and GRE sequence designs may render laser discectomies more effective and controllable.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app