Clinical performance of infrared consumer-grade thermometers.

BACKGROUND AND PURPOSE: Technology of ear infrared (IR) thermometers has improved. This study compared a modern ear thermometer to forehead or temporal artery thermometers.

METHODS: Temperatures were measured with a heated-tip ear thermometer, a temporal artery thermometer, 3 forehead thermometers, and a thermistor-based reference thermometer in monitor mode.

RESULTS: In 171 subjects, mean bias with the forehead thermometers was significantly higher (p < .001) than with the ear thermometer (0.01 degrees C +/- 0.41 degrees C). In 64 febrile subjects, bias with the ear thermometer was significantly lower than with 3 of the other thermometers. A false-negative reading was less likely with the ear thermometer (8%) versus the others (55%, 56%, 28%, and 47%).

CONCLUSIONS: Modem ear thermometry provides more precise measurements closer to those of a reference thermometer and is less likely to give false-negative readings than forehead or temporal artery measurements.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app