Systematic fluoroscopic and electrical assessment of implantable cardioverter-defibrillator patients implanted with silicone-polyurethane copolymer (Optim™) coated leads.

AIMS: Serious concerns have been recently raised about the reliability of the silicone-polyurethane copolymer (Optim™) lead insulation system. We sought to identify insulation defects and Optim-lead failures by systematic fluoroscopic and electrical assessment in a prospectively defined cohort of implantable cardioverter-defibrillator (ICD) patients.

METHODS AND RESULTS: Between July 2007 and December 2011, 234 patients were implanted with 413 optim-coated leads as part of an ICD system at a single centre. Fluoroscopic screening with high-resolution cine-fluoroscopy at 30 frames per second was offered to all patients. In addition, the electrical integrity of all implanted leads was assessed. Durata, Riata ST Optim, and low-voltage Optim leads were implanted in 199, 26, and 188 cases, respectively. During a total follow-up of 10 036 lead-months, there were 7 Optim-lead failures (defined as electrical malfunction resulting in lead replacement) and 31 deaths; no cases of electrical noises were encountered. The overall incidence of lead failure was 1.2 vs. 0.3 per 100 lead-years, for high- and low-voltage leads, respectively (P = 0.1). One hundred fifty-one patients agreed to undergo fluoroscopy screening; none of the 264 analysed Optim leads were found to have any fluoroscopically visible structural defects after an average of 31 months post-implant.

CONCLUSION: This study represents the first systematic screening of Optim-coated leads in a large unselected cohort of ICD patients. Over a 5-year period few lead failures were observed and normal fluoroscopic appearance was present in all patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app