Comparison of the actual amount of axial movement of 3 aspheric intraocular lenses using anterior segment optical coherence tomography.
Journal of Cataract and Refractive Surgery 2013 October
PURPOSE: To compare the mean effective lens position (ELP) and actual amount of axial movement of 3 aspheric intraocular lenses (IOLs) using anterior segment optical coherence tomography (AS-OCT).
SETTING: Department of Ophthalmology, Korea University College of Medicine, Seoul, South Korea.
DESIGN: Retrospective comparative study.
METHODS: Consecutive patients had phacoemulsification with aspheric IOL implantation. The ELP measurements by AS-OCT were taken 1 week and 1, 3, and 6 months postoperatively. The ELPRMS was defined as the root mean square (RMS) of the change in ELP at each follow-up timepoint.
RESULTS: An XL Stabi ZO 3-plate IOL was implanted in 30 eyes, an Acrysof IQ C-loop IOL in 22 eyes, and an Akreos MI-60 4-plate IOL in 17 eyes. The 4-plate IOL showed significant changes in the mean ELP from 1 week to 1 month and from 3 to 6 months postoperatively (P=.005 and P=.028, respectively). The changes of the other 2 IOLs were insignificant during the 3 postoperative visits. However, the mean ELPRMS of the C-loop IOL (0.06 mm ± 0.31 [SD]) was smaller than that of the 3-plate IOL (0.14 ± 0.28 mm) and the 4-plate IOL (0.20 ± 0.35 mm) (P=.014 and P=.023, respectively) during the 3 timepoints.
CONCLUSION: An appropriate method based on the actual amount of axial IOL movement, such as the ELPRMS, is needed to assess the axial positional stability of IOLs.
FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
SETTING: Department of Ophthalmology, Korea University College of Medicine, Seoul, South Korea.
DESIGN: Retrospective comparative study.
METHODS: Consecutive patients had phacoemulsification with aspheric IOL implantation. The ELP measurements by AS-OCT were taken 1 week and 1, 3, and 6 months postoperatively. The ELPRMS was defined as the root mean square (RMS) of the change in ELP at each follow-up timepoint.
RESULTS: An XL Stabi ZO 3-plate IOL was implanted in 30 eyes, an Acrysof IQ C-loop IOL in 22 eyes, and an Akreos MI-60 4-plate IOL in 17 eyes. The 4-plate IOL showed significant changes in the mean ELP from 1 week to 1 month and from 3 to 6 months postoperatively (P=.005 and P=.028, respectively). The changes of the other 2 IOLs were insignificant during the 3 postoperative visits. However, the mean ELPRMS of the C-loop IOL (0.06 mm ± 0.31 [SD]) was smaller than that of the 3-plate IOL (0.14 ± 0.28 mm) and the 4-plate IOL (0.20 ± 0.35 mm) (P=.014 and P=.023, respectively) during the 3 timepoints.
CONCLUSION: An appropriate method based on the actual amount of axial IOL movement, such as the ELPRMS, is needed to assess the axial positional stability of IOLs.
FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app