One-year hemodynamic comparison of Perimount Magna with St Jude Epic aortic bioprostheses.
Archives of Medical Science : AMS 2013 June 21
INTRODUCTION: Cardiac surgeons are using more bioprosthetic valves due to the ageing population as well as to improvements that have been made to these implants. We sought to compare the 1-year hemodynamics of two commercially available valves by echocardiographic parameters.
MATERIAL AND METHODS: Retrospective review of our institutional database revealed 69 patients who received either Perimount Magna (n = 33) or St Jude Epic (n = 36) valves in the aortic position with no other valve surgery between June 2004 and March 2006. All patients received transthoracic echocardiography at 1 year. Comparisons between groups were made at baseline and at 1-year follow-up. In addition, a pairwise comparison was performed in each patient to determine the change in echocardiographic parameters between baseline and follow-up.
RESULTS: Mean implanted valve size was similar (Magna 24.3 ±2.0 mm vs. Epic 24.1 ±2.2 mm). Pre- and intraoperative patient variables were similar between the two groups. There were lower peak and mean pressure gradients in the Magna group, both at discharge and one year after surgery. This correlated with a larger indexed effective orifice area (Magna 0.8 ±0.2 cm(2)/m(2) vs. Epic 0.67 ±0.2 cm(2)/m(2), p = 0.02). In spite of these findings, left ventricular mass regression was not different.
CONCLUSIONS: These findings suggest that in a series with relatively low indexed effective orifice areas, the peak and mean gradients obtained were acceptable. More clinical follow-up of these patients is required to assess the true impact of prosthesis patient mismatch.
MATERIAL AND METHODS: Retrospective review of our institutional database revealed 69 patients who received either Perimount Magna (n = 33) or St Jude Epic (n = 36) valves in the aortic position with no other valve surgery between June 2004 and March 2006. All patients received transthoracic echocardiography at 1 year. Comparisons between groups were made at baseline and at 1-year follow-up. In addition, a pairwise comparison was performed in each patient to determine the change in echocardiographic parameters between baseline and follow-up.
RESULTS: Mean implanted valve size was similar (Magna 24.3 ±2.0 mm vs. Epic 24.1 ±2.2 mm). Pre- and intraoperative patient variables were similar between the two groups. There were lower peak and mean pressure gradients in the Magna group, both at discharge and one year after surgery. This correlated with a larger indexed effective orifice area (Magna 0.8 ±0.2 cm(2)/m(2) vs. Epic 0.67 ±0.2 cm(2)/m(2), p = 0.02). In spite of these findings, left ventricular mass regression was not different.
CONCLUSIONS: These findings suggest that in a series with relatively low indexed effective orifice areas, the peak and mean gradients obtained were acceptable. More clinical follow-up of these patients is required to assess the true impact of prosthesis patient mismatch.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app