Intermittent subglottic secretion drainage may cause tracheal damage in patients with few oropharyngeal secretions.
Intensive & Critical Care Nursing : the Official Journal of the British Association of Critical Care Nurses 2013 December
OBJECTIVE: Injurious prolapse of tracheal mucosa into the suction port has been reported in up to 50% of intubated patients receiving continuous aspiration of subglottic secretions. We investigated whether similar injury could be inflicted by automated intermittent aspiration.
METHODS: Six consecutive patients, intubated with the Mallinckrodt TaperGuard Evac™ endotracheal tube, were studied. A flow sensor was placed between the vacuum regulating system and the mucus collector. Intermittent suctioning was performed at a pressure of -125 mmHg with a 25s interval and duration of 15s. After 24h, a CT scan of the tracheal region was performed.
RESULTS: Excessive negative suction pressure, a fast drop in aspiration flow to zero, and important "swinging" movements of secretions in the evacuation line were observed in all patients. Oral instillation of antiseptic mouthwash restored normal aspiration flow and secretion mobility. CT imaging showed marked entrapment of tracheal mucosa into the suction port in all patients.
CONCLUSION: In patients with few oropharyngeal secretions, automated intermittent subglottic aspiration may result in significant and potential harmful invagination of tracheal mucosa into the suction lumen. A critical amount of fluid must be present in the oropharynx to assure adequate and safe aspiration.
METHODS: Six consecutive patients, intubated with the Mallinckrodt TaperGuard Evac™ endotracheal tube, were studied. A flow sensor was placed between the vacuum regulating system and the mucus collector. Intermittent suctioning was performed at a pressure of -125 mmHg with a 25s interval and duration of 15s. After 24h, a CT scan of the tracheal region was performed.
RESULTS: Excessive negative suction pressure, a fast drop in aspiration flow to zero, and important "swinging" movements of secretions in the evacuation line were observed in all patients. Oral instillation of antiseptic mouthwash restored normal aspiration flow and secretion mobility. CT imaging showed marked entrapment of tracheal mucosa into the suction port in all patients.
CONCLUSION: In patients with few oropharyngeal secretions, automated intermittent subglottic aspiration may result in significant and potential harmful invagination of tracheal mucosa into the suction lumen. A critical amount of fluid must be present in the oropharynx to assure adequate and safe aspiration.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app