PURPOSE: To evaluate the effect of reducing the number of visual acuity measurements made in a defocus curve on the quality of data quantified.

SETTING: Midland Eye, Solihull, United Kingdom.

DESIGN: Evaluation of a technique.

METHODS: Defocus curves were constructed by measuring visual acuity on a distance logMAR letter chart, randomizing the test letters between lens presentations. The lens powers evaluated ranged between +1.50 diopters (D) and -5.00 D in 0.50 D steps, which were also presented in a randomized order. Defocus curves were measured binocularly with the Tecnis diffractive, Rezoom refractive, Lentis rotationally asymmetric segmented (+3.00 D addition [add]), and Finevision trifocal multifocal intraocular lenses (IOLs) implanted bilaterally, and also for the diffractive IOL and refractive or rotationally asymmetric segmented (+3.00 D and +1.50 D adds) multifocal IOLs implanted contralaterally. Relative and absolute range of clear-focus metrics and area metrics were calculated for curves fitted using 0.50 D, 1.00 D, and 1.50 D steps and a near add-specific profile (ie, distance, half the near add, and the full near-add powers).

RESULTS: A significant difference in simulated results was found in at least 1 of the relative or absolute range of clear-focus or area metrics for each of the multifocal designs examined when the defocus-curve step size was increased (P<.05).

CONCLUSION: Faster methods of capturing defocus curves from multifocal IOL designs appear to distort the metric results and are therefore not valid.

FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app