Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients.

BACKGROUND: A new calibrated pulse wave analysis method (VolumeView™/EV1000™, Edwards Lifesciences, Irvine, CA, USA) has been developed to continuously monitor cardiac output (CO). The aim of this study was to compare the performance of the VolumeView method, and of the PiCCO2™ pulse contour method (Pulsion Medical Systems, Munich, Germany), with reference transpulmonary thermodilution (TPTD) CO measurements.

METHODS: This was a prospective, multicentre observational study performed in the surgical and interdisciplinary intensive care units of four tertiary hospitals. Seventy-two critically ill patients were monitored with a central venous catheter, and a thermistor-tipped femoral arterial VolumeView™ catheter connected to the EV1000™ monitor. After initial calibration by TPTD CO was continuously assessed using the VolumeView-CCO software (CCO(VolumeView)) during a 72 h period. TPTD was performed in order to obtain reference CO values (COREF). TPTD and arterial wave signals were transmitted to a PiCCO2™ monitor in order to obtain CCO(PiCCO) values. CCO(VolumeView) and CCO(PiCCO) were recorded over a 5 min interval before assessment of CO(TPTD). Bland-Altman analysis, %(errors), and concordance (trend analysis) were calculated.

RESULTS: A total of 338 matched sets of data were available for comparison. Bias for CCO(VolumeView)-CO(REF) was -0.07 litre min(-1) and for CCO(PiCCO)-CO(REF) +0.03 litre min(-1). Corresponding limits of agreement were 2.00 and 2.48 litre min(-1) (P<0.01), %(errors) 29 and 37%, respectively. Trending capabilities were comparable for both techniques.

CONCLUSIONS: The performance of the new VolumeView™-CCO method is as reliable as the PiCCO2™-CCO pulse wave analysis in critically ill patients. However, an improved precision was observed with the VolumeView™ technique. CLINICALTRIALS.GOV IDENTIFIER: NCT01405040.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app