Safety of repetitive nerve stimulation in patients with cardiac implantable electronic devices.
Muscle & Nerve 2013 June
INTRODUCTION: Pacemakers and implantable cardioverter-defibrillators (ICDs) are vulnerable to inappropriate sensing of electromagnetic interference (EMI), such as from nerve conduction studies. We conducted a prospective study to assess the safety of repetitive nerve stimulation (RNS).
METHODS: Fourteen patients undergoing insertion of 10 ICDs and 4 pacemakers under general anesthesia received RNS of the median, axillary, and spinal accessory nerves at 2 HZ and 50 HZ. We recorded detection of EMI and whether or not this resulted in an arrhythmia diagnosis or change in pacing output.
RESULTS: EMI was visible in 2 ICDs, without spurious tachyarrhythmia detection. EMI was observed in 3 of the 4 pacemakers, which led to pacing inhibition and a pause in 2, both of which were programmed to a unipolar sensing configuration.
CONCLUSIONS: RNS is safe in patients with ICDs. In pacemaker patients, RNS appears safe during bipolar sensing, but caution is recommended in pacemaker patients with unipolar sensing.
METHODS: Fourteen patients undergoing insertion of 10 ICDs and 4 pacemakers under general anesthesia received RNS of the median, axillary, and spinal accessory nerves at 2 HZ and 50 HZ. We recorded detection of EMI and whether or not this resulted in an arrhythmia diagnosis or change in pacing output.
RESULTS: EMI was visible in 2 ICDs, without spurious tachyarrhythmia detection. EMI was observed in 3 of the 4 pacemakers, which led to pacing inhibition and a pause in 2, both of which were programmed to a unipolar sensing configuration.
CONCLUSIONS: RNS is safe in patients with ICDs. In pacemaker patients, RNS appears safe during bipolar sensing, but caution is recommended in pacemaker patients with unipolar sensing.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app