COMPARATIVE STUDY
JOURNAL ARTICLE
MULTICENTER STUDY
RANDOMIZED CONTROLLED TRIAL
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study.

BACKGROUND: Pulmonary vein isolation is the most prevalent approach for catheter ablation of paroxysmal atrial fibrillation. Long-term success of the procedure is diminished by arrhythmia recurrences occurring predominantly because of reconnections in previously isolated pulmonary veins. The aim of the EFFICAS I multicenter study was to demonstrate the correlation between contact force (CF) parameters during initial procedure and the incidence of isolation gaps (gap) at 3-month follow-up.

METHOD AND RESULTS: A radiofrequency ablation catheter with integrated CF sensor (TactiCath, Endosense, Geneva, Switzerland) was used to perform pulmonary vein isolation in 46 patients with paroxysmal atrial fibrillation. During the ablation procedure, the operator was blinded to CF information. At follow-up, an interventional diagnostic procedure was performed to assess gap location as correlated to index procedure ablation parameters. At follow-up, 65% (26/40) of patients showed ≥1 gaps. Ablations with minimum Force-Time Integral (FTI) <400 gs showed increased likelihood for reconnection (P<0.001). Reconnection correlated strongly with minimum CF (P<0.0001) and minimum FTI (P=0.0007) at the site of gap. Gap occurrence showed a strong trend with lower average CF and average FTI. CF and FTI are generally higher on the right side, although the left anterior segment presents a unique challenge to achieve stable position with good CF.

CONCLUSIONS: Minimum CF and minimum FTI values are strong predictors of gap formation. Optimal CF parameter recommendations are a target CF of 20 g and a minimum FTI of 400 gs for each new lesion.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app