Model-based prediction of portal dose images during patient treatment.

Medical Physics 2013 March
PURPOSE: Dosimetric verification of radiation therapy is crucial when delivering complex treatments like intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy (VMAT. Pretreatment verification, characterized by methods applied without the patient present and before the treatment start date, is typically carried out at most centers. In vivo dosimetric verification, characterized by methods applied with the patient present, is not commonly carried out in the clinic. This work presents a novel, model-based EPID dosimetry method that could be used for routine clinical in vivo patient treatment verification.

METHODS: The authors integrated a detailed fluence model with a patient scatter prediction model that uses a superposition of scatter energy fluence kernels, generated via Monte Carlo techniques, to determine patient scatter fluence delivered to the EPID. The total dose to the EPID was calculated using the sum of convolutions of the calculated energy fluence distribution entering the EPID with monoenergetic dose kernels, specific to the a-Si EPID. Measured images with simple, square fields delivered to slab phantoms were validated against predicted images. Measured and predicted images acquired during the delivery of IMRT fields to slabs and an anthropomorphic phantom were compared using the χ-comparison for 3% dose difference and 3 mm distance-to-agreement criteria.

RESULTS: Predicted and measured images of the square fields with slabs in the field agreed within 2.5%. Predicted portal dose images of clinical IMRT fields delivered to slabs and an anthropomorphic phantom agreed with measured images within 3% and 3 mm for an average of at least 97% of the infield pixels (defined as >10% maximum field dose) for each case, over all fields.

CONCLUSIONS: This work presents the first validation of the integration of a comprehensive fluence model with a patient and EPID radiation transport model that accounts for patient transmission, including complex factors such as patient scatter and the energy response of the a-Si detector. The portal dose image prediction model satisfies the 3% and 3 mm criteria for IMRT fields delivered to slab phantoms and could be used for patient treatment verification.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app