Percutaneous interventions on the hemodialysis reliable outflow vascular access device.
PURPOSE: To determine the outcomes of percutaneous interventions for prolonging the patency of the Hemodialysis Reliable Outflow (HeRO) device.
MATERIALS AND METHODS: Between January 2007 and August 2011, 73 percutaneous interventions were performed on 26 HeRO devices in 25 patients. The graft was implanted in the upper arm with the outflow catheter tip in the superior vena cava or right atrium. Procedural reports, angiographic images, and clinical notes were retrospectively reviewed. The primary and secondary patency rates after intervention were calculated using the Kaplan-Meier method.
RESULTS: The mean time from HeRO implantation to initial dysfunction or thrombosis was 171 days. In 60 (82%) procedures, the HeRO device was thrombosed. An intragraft stenosis was the most common lesion identified (59%; n = 43) followed by an arterial anastomosis stenosis identified in 18% (n = 13). In 22% (n = 16) of procedures in which the HeRO device was thrombosed, an underlying cause was not identified after thrombectomy. The 3-, 6-, and 12-month primary patency rates after intervention were 47%, 37%, and 26% for first-time interventions. The secondary patency rates were 80%, 70%, and 64%. The only complication was pulmonary embolism resulting in death 2 days after HeRO thrombectomy.
CONCLUSIONS: Percutaneous interventions on thrombosed and failing HeRO devices yielded acceptable primary and secondary patency rates after intervention in these patients with few, if any, alternatives for hemodialysis access.
MATERIALS AND METHODS: Between January 2007 and August 2011, 73 percutaneous interventions were performed on 26 HeRO devices in 25 patients. The graft was implanted in the upper arm with the outflow catheter tip in the superior vena cava or right atrium. Procedural reports, angiographic images, and clinical notes were retrospectively reviewed. The primary and secondary patency rates after intervention were calculated using the Kaplan-Meier method.
RESULTS: The mean time from HeRO implantation to initial dysfunction or thrombosis was 171 days. In 60 (82%) procedures, the HeRO device was thrombosed. An intragraft stenosis was the most common lesion identified (59%; n = 43) followed by an arterial anastomosis stenosis identified in 18% (n = 13). In 22% (n = 16) of procedures in which the HeRO device was thrombosed, an underlying cause was not identified after thrombectomy. The 3-, 6-, and 12-month primary patency rates after intervention were 47%, 37%, and 26% for first-time interventions. The secondary patency rates were 80%, 70%, and 64%. The only complication was pulmonary embolism resulting in death 2 days after HeRO thrombectomy.
CONCLUSIONS: Percutaneous interventions on thrombosed and failing HeRO devices yielded acceptable primary and secondary patency rates after intervention in these patients with few, if any, alternatives for hemodialysis access.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app