Feasibility of high-dose dobutamine stress SSFP Cine MRI at 3 Tesla with patient adaptive local RF Shimming using dual-source RF transmission: initial results.
PURPOSE: To investigate the feasibility of high-dose dobutamine stress (HDDS) imaging using SSFP sequences at 3 T employing patient-adaptive local RF-shimming using a dual-source RF transmission system.
MATERIALS AND METHODS: 13 Patients underwent a HDDS protocol on a 3 T MRI scanner (Achieva 3.0T-TX, Philips Healthcare), equipped with a dual-source RF transmission system. SSFP cine sequences using patient-adaptive local RF-shimming (RF-S) were compared to cine images acquired without additional shimming. Image quality was evaluated on a 4-point grading scale and number of non-diagnostic segments assessed. Contrast (CN) between myocardium (SIM) and blood pool (SIB) was calculated [(SIB-SIM)/(SIB+SIM)].
RESULTS: Image quality both at rest and maximum stress was significantly improved with RF-S (ED:3.56±0.5 vs. 3.23±0.63; ES:3.4±0.5 vs. 3.1±0.7) compared to no RF-S (ED:2.9±0.72 vs. 2.15±0.78; ES:2.64±0.74 vs. 1.95±0.76; p<0.01). The amount of non-diagnostic segments was significantly reduced when using RF-S at rest and stress (3 vs. 39; 19 vs. 78, p<0.05). All HDDS studies were diagnostic if performed with RF-S (n=13/13) in comparison to conventional shimming (n=5/13). Image contrast was improved for SSFP sequences with RF-S (0.53±0.08) compared to conventional images (0.46±0.09, p=0.06).
CONCLUSION: Patient-adaptive local RF-shimming using a dual-source RF transmission system allows for reliable SSFP imaging in a clinical high-dose dobutamine stress protocol at 3 T. RF-S significantly improves image quality and reduces the number of non-diagnostic myocardial segments.
MATERIALS AND METHODS: 13 Patients underwent a HDDS protocol on a 3 T MRI scanner (Achieva 3.0T-TX, Philips Healthcare), equipped with a dual-source RF transmission system. SSFP cine sequences using patient-adaptive local RF-shimming (RF-S) were compared to cine images acquired without additional shimming. Image quality was evaluated on a 4-point grading scale and number of non-diagnostic segments assessed. Contrast (CN) between myocardium (SIM) and blood pool (SIB) was calculated [(SIB-SIM)/(SIB+SIM)].
RESULTS: Image quality both at rest and maximum stress was significantly improved with RF-S (ED:3.56±0.5 vs. 3.23±0.63; ES:3.4±0.5 vs. 3.1±0.7) compared to no RF-S (ED:2.9±0.72 vs. 2.15±0.78; ES:2.64±0.74 vs. 1.95±0.76; p<0.01). The amount of non-diagnostic segments was significantly reduced when using RF-S at rest and stress (3 vs. 39; 19 vs. 78, p<0.05). All HDDS studies were diagnostic if performed with RF-S (n=13/13) in comparison to conventional shimming (n=5/13). Image contrast was improved for SSFP sequences with RF-S (0.53±0.08) compared to conventional images (0.46±0.09, p=0.06).
CONCLUSION: Patient-adaptive local RF-shimming using a dual-source RF transmission system allows for reliable SSFP imaging in a clinical high-dose dobutamine stress protocol at 3 T. RF-S significantly improves image quality and reduces the number of non-diagnostic myocardial segments.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app