Stability of electrolyte determinations on the Siemens Advia 1800 analyzer.

BACKGROUND: It is sometimes necessary for the laboratory to re-test samples for critical serum electrolyte levels. It is important to assure reproducibility of results when testing is performed on stored, refrigerated samples. We have tested the reproducibility of results for the critical electrolytes, Na, K, Cl and Ca, from ten randomly selected patients'sera over our maximum storage period of nine (9) days on the Siemens Advia 1800 analyzer. The ranges for each electrolyte were 131-150 meq/L (Na), 3.4-5.2 meq/L (K), 101-123 meq/L (Cl) and 7.3-9.9 mg/dL (Ca).

METHODS: We used ion-selective electrodes for Na, K and Cl and the ortho-cresolphthalein dye method for Ca.

RESULTS: We find that the reproducibility of determinations for all of these electrolytes was excellent, i.e. the coefficients of variation for each electrolyte determination for each patient were low.

CONCLUSION: The methods of measurement for these electrolytes on the Advia 1800 are reliable and reproducible.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app