Role of clinician's experience and implant design on implant stability. An ex vivo study in artificial soft bones.

OBJECTIVES: Clinical experience in implant placement is important in order to prevent implant failures. However, the implant design affects the primary implant stability (PS) especially in poor quality bones. Therefore, the aim of this study was to compare the effect of clinician surgical experience on PS, when placing different type of implant designs.

METHODS: A total of 180 implants (90 parallel walled-P and 90 tapered-T) were placed in freshly slaughtered cow ribs. Bone quality was evaluated by two examiners during surgery and considered as 'type IV' bone. Implants (ø 5 mm, length: 15 mm, Osseotite, BIOMET 3i, Palm Beach Gardens, FL, USA) were placed by three different clinicians (master/I, good/II, non-experienced/III, under direct supervision of a manufacturer representative; 30 implants/group). An independent observer assessed the accuracy of placement by resonance frequency analysis (RFA) with implant stability quotient (ISQ) values. Two-way analysis of variance (ANOVA) and Tukey's post hoc test were used to detect the surgical experience of the clinicians and their interaction and effects of implant design on the PS.

RESULTS: All implants were mechanically stable. The mean ISQ values were: 49.57(± 18.49) for the P-implants and 67.07(± 8.79) for the T-implants. The two-way ANOVA showed significant effects of implant design (p < .0001), clinician (p < .0001), and their interaction (p < .0001). The Tukey's multiple comparison test showed significant differences in RFA for the clinician group I/II (p = .015) and highly significant (p < .0001) between I/III and II/III. The P-implants presented (for I, II, and III) mean ISQ values 31.25/49.18/68.17 and the T-implants showed higher ISQ values, 70.15/62.08/68.98, respectively. Clinicians I and II did not show extreme differences for T-implants (p = .016). In contrast, clinician III achieved high ISQ values using P- and T-implants following the exact surgical protocol based on the manufacturer guidelines. T-implants provided high stability for experienced clinicians compared with P-implants.

CONCLUSION: T-implants achieved greater PS than the P-implants. All clinicians consistently achieved PS; however, experienced clinicians achieved higher ISQ values with T-implants in poor quality bone.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app