A biomechanical comparison of static versus dynamic lag screw modes for cephalomedullary nails used to fix unstable peritrochanteric fractures.
Journal of Trauma and Acute Care Surgery 2012 Februrary
BACKGROUND: The gamma nail has an option to statically lock its lag screw (static mode) or to allow its lag screw to move within the nail to compress the intertrochanteric fracture (dynamic mode). The purpose of this study was to compare the biomechanical stiffness of static and dynamic lag screw modes for a cephalomedullary nail used to fix an unstable peritrochanteric fracture.
METHODS: Unstable four-part peritrochanteric fractures were created in 30 synthetic femurs and fixed with Long Gamma 3 Nails. Mechanical tests were conducted for axial, lateral, and torsional stiffness with intact femurs, femur-nail constructs with static lag screw mode,and femur-nail constructs with dynamic lag screw mode. A paired Student's t test was used for all statistical comparisons between test groups.
RESULTS: Axial and torsional stiffness of intact femurs was significantly greater than femur-nail constructs (p < 0.01 all comparisons),whereas lateral stiffness was significantly less (p < 0.01 all comparisons). Axial stiffness of the femur-nail construct was significantly greater (p < 0.01) in static mode (484.3 N/mm 80.2 N/mm) than in dynamic mode (424.1 N/mm 78.0 N/mm).Lateral stiffness was significantly greater (p < 0.01) in static mode (113.9 N/mm 8.4 N/mm) than in dynamic mode (109.5N/mm 8.8 N/mm). Torsional stiffness was significantly greater (p = 0.02) in dynamic mode (114.5 N/mm 28.2 N/mm) than in static mode (111.7 N/mm 27.0 N/mm).
DISCUSSION: There is a 60 N/mm (12.4%) reduction in axial stiffness when the lag screw is in dynamic mode. Given the statistically significant reduction in axial and lateral stiffness with use of the dynamic mode, static lag screw mode should be further explored clinically for treatment of unstable peritrochanteric fractures.
METHODS: Unstable four-part peritrochanteric fractures were created in 30 synthetic femurs and fixed with Long Gamma 3 Nails. Mechanical tests were conducted for axial, lateral, and torsional stiffness with intact femurs, femur-nail constructs with static lag screw mode,and femur-nail constructs with dynamic lag screw mode. A paired Student's t test was used for all statistical comparisons between test groups.
RESULTS: Axial and torsional stiffness of intact femurs was significantly greater than femur-nail constructs (p < 0.01 all comparisons),whereas lateral stiffness was significantly less (p < 0.01 all comparisons). Axial stiffness of the femur-nail construct was significantly greater (p < 0.01) in static mode (484.3 N/mm 80.2 N/mm) than in dynamic mode (424.1 N/mm 78.0 N/mm).Lateral stiffness was significantly greater (p < 0.01) in static mode (113.9 N/mm 8.4 N/mm) than in dynamic mode (109.5N/mm 8.8 N/mm). Torsional stiffness was significantly greater (p = 0.02) in dynamic mode (114.5 N/mm 28.2 N/mm) than in static mode (111.7 N/mm 27.0 N/mm).
DISCUSSION: There is a 60 N/mm (12.4%) reduction in axial stiffness when the lag screw is in dynamic mode. Given the statistically significant reduction in axial and lateral stiffness with use of the dynamic mode, static lag screw mode should be further explored clinically for treatment of unstable peritrochanteric fractures.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app