Development and validation of rapid use scope test strips to determine the efficacy of manual cleaning for flexible endoscope channels.
American Journal of Infection Control 2012 November
BACKGROUND: Cleaning of flexible endoscopes is most commonly performed using manual methods that are often performed inadequately. The aim of this study was to validate the sample collection protocol and the Rapid Use Scope Test (RUST) and then assess its usefulness in clinical use.
METHODS: The benchmarks for adequate cleaning were protein <6.4 μg/cm(2), hemoglobin <2.2 μg/cm(2), and carbohydrate <1.2 μg/cm(2). Sample collection consisted of flushing 10 mL of sterile reverse osmosis water through the suction-biopsy port to the distal end. Validation of the RUST audit tool included simulated-use and in-use testing in 43 endoscopy clinics across Canada.
RESULTS: Simulated-use testing validated that improperly cleaned endoscopes that exceeded the cleaning benchmarks would be flagged by the RUST test. The clinical-use study indicated that 96.6% of 1,489 scope channels tested were RUST negative; however, 19% and 12% of elevator guide-wire channels and endoscopic retrograde colangiopancreatography channels, respectively, exceeded the benchmarks. The survey indicated that reprocessing personnel valued a rapid audit tool for assessing compliance with manual cleaning.
CONCLUSION: The validated RUST test provides health care users with a rapid audit tool for manual cleaning that can be integrated into the quality program in endoscopy.
METHODS: The benchmarks for adequate cleaning were protein <6.4 μg/cm(2), hemoglobin <2.2 μg/cm(2), and carbohydrate <1.2 μg/cm(2). Sample collection consisted of flushing 10 mL of sterile reverse osmosis water through the suction-biopsy port to the distal end. Validation of the RUST audit tool included simulated-use and in-use testing in 43 endoscopy clinics across Canada.
RESULTS: Simulated-use testing validated that improperly cleaned endoscopes that exceeded the cleaning benchmarks would be flagged by the RUST test. The clinical-use study indicated that 96.6% of 1,489 scope channels tested were RUST negative; however, 19% and 12% of elevator guide-wire channels and endoscopic retrograde colangiopancreatography channels, respectively, exceeded the benchmarks. The survey indicated that reprocessing personnel valued a rapid audit tool for assessing compliance with manual cleaning.
CONCLUSION: The validated RUST test provides health care users with a rapid audit tool for manual cleaning that can be integrated into the quality program in endoscopy.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app