We have located links that may give you full text access.
Comparative Study
Journal Article
Multicenter Study
Randomized Controlled Trial
Research Support, Non-U.S. Gov't
Reduction of shunt obstructions by using a peel-away sheath technique? A multicenter prospective randomized trial.
Clinical Neurology and Neurosurgery 2012 May
OBJECTIVE: Shunt obstructions may partly be caused by brain debris, which intrude into the ventricular catheter during ventricle puncture. Avoiding contact between the catheter and brain tissue, by using a peel-away sheath, should reduce the number of shunt failures caused by obstruction. To test this hypothesis, we conducted a randomized, prospective multicenter study.
METHODS: 201 patients from 6 different neurosurgical centers in Germany receiving a ventriculo-peritoneal shunt were included in this study. Of these, 177 patients completed a 1-year follow-up period. Surgery was randomized in a 1 to 1 fashion, such that out of 177 procedures, 91 were performed using a peel-away sheath and 86 were performed without. The rate of surgical re-interventions and shunt obstructions within a 12-month period was recorded.
RESULTS: Within 1 year post-surgery, 17 shunt obstructions (9.6%) leading to shunt revisions were recorded. However, no difference was found between surgeries performed using a peel-away sheath (9.9%) or not (9.3%). The overall shunt infection rate was 2.8% and the shunt revision rate for overdrainage was 3.9%.
CONCLUSIONS: The theoretical advantages attributed to the use of a peel-away sheath to introduce a ventricular catheter could not be confirmed in this randomized study, suggesting that the proposed role of brain debris in shunt obstructions may be overestimated.
METHODS: 201 patients from 6 different neurosurgical centers in Germany receiving a ventriculo-peritoneal shunt were included in this study. Of these, 177 patients completed a 1-year follow-up period. Surgery was randomized in a 1 to 1 fashion, such that out of 177 procedures, 91 were performed using a peel-away sheath and 86 were performed without. The rate of surgical re-interventions and shunt obstructions within a 12-month period was recorded.
RESULTS: Within 1 year post-surgery, 17 shunt obstructions (9.6%) leading to shunt revisions were recorded. However, no difference was found between surgeries performed using a peel-away sheath (9.9%) or not (9.3%). The overall shunt infection rate was 2.8% and the shunt revision rate for overdrainage was 3.9%.
CONCLUSIONS: The theoretical advantages attributed to the use of a peel-away sheath to introduce a ventricular catheter could not be confirmed in this randomized study, suggesting that the proposed role of brain debris in shunt obstructions may be overestimated.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app