Analysis of complications of permanent transvenous implantable cardiac pacemaker related to operative and postoperative management in 717 consecutive patients.
Hiroshima Journal of Medical Sciences 1990 December
A retrospective study on complications especially related to operative or postoperative management was carried out in 1000 pacemaker implantations in 717 patients between September, 1972 and December, 1986. In 33 of our series (4.8%), 24 pacing failure occurred within two weeks of implantation. Flange-type leads had the highest incidence of complications. Wedging the electrode close to the longitudinal axis of the heart was found to be the best placing, assuming that the tip of the electrode and the section immediately adjacent were pointing downward. There were 6 cases of diaphragmatic pacing which could not be corrected through programming. No relation with the position of the electrode could be found. We recommend using bipolar pacing to lower the incidence of diaphragmatic pacing. We also had 9 patients with lead fractures (1.3%); these included 3 cases with silicone insulation breaks and 6 cases with wire fractures which occurred closely proximal to the area where the lead was fixed to the vessels. No relationship between the ratio of lead fractures and their brand was found. Five patients developed infections, all of them late postoperatively, over a period of 8 months to 5 years postop. Of these, there were 3 cases with postoperative prolonged hematoma at the pocket site, and 4 cases which had required lead repositioning because of pacing failure. The incidence of infection in our series was low when compared to previous reports, probably due to local use of one gram of Kanamycin during the operation and active chemotherapy performed short term postoperatively. To manage infection of the skin pocket, the pacemaker was removed and a new pacemaker was implanted in the opposite side. There was 7 early postoperative deaths. One of them due to cardiac tamponade caused by perforation of the cardiac wall by temporary electrode lead. The resustation was unsuccessful.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app