Flow-through versus sidestream capnometry for detection of end tidal carbon dioxide in the sedated patient.

BACKGROUND: End tidal carbon dioxide (ETCO(2)) in non-intubated patients can be monitored using either sidestream or flow-through capnometry [Yamamori et al., J Clin Monit Comput 22(3):209-220, 2008]. The hypothesis of this validation study is that, flow-through capnometry will yield a more accurate estimate of ETCO(2) than sidestream capnometry when evaluated in a bench study during low tidal volumes and high oxygen administration via nasal cannula. Secondarily, when ETCO(2) from each is compared to arterial CO(2) (PaCO(2)) during a study in which healthy, non-intubated volunteers are tested under normocapnic, hypocapnic and hypercapnic conditions, the flow-through capnometer will resemble PaCO(2) more closely than the sidestream capnometer. This will be especially true during periods of lower minute ventilation and high oxygen flow rates via mask in non-intubated, remifentanil sedated, healthy volunteers whose physiologic deadspace is small.

METHODS: The performance of a flow-through (cap-ONE, Nihon Kohden, Tokyo, Japan) and a sidestream (Microcap Smart CapnoLine Plus, Oridion Inc., Needham, MA) capnometer were compared in a bench study and a volunteer trial. A bench study evaluated ETCO(2) accuracy using waveforms generated via mechanical lungs during low tidal volumes and high oxygen flow rates. A volunteer study compared the ETCO(2) for each capnometer against PaCO(2) during sedation in which 8 l O(2) was delivered via mask rather than the nasal cannula.

RESULTS: In the bench study, the flow-through capnometer gave slightly higher values of ETCO(2) during high-flow oxygen and no discernable differences during variable tidal volumes. Bland and Altman plots comparing ETCO(2) to PaCO(2) showed essentially equal performance between the two capnometers in the volunteers.

CONCLUSIONS: Within a wide limit of agreement between the volunteer and bench study, flow-through and sidestream capnometry performed equally well during bench testing and in non-intubated, sedated patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app