Microbiological challenge of four protective devices for the reconstitution of cytotoxic agents.
Letters in Applied Microbiology 2008 December
AIMS: To evaluate the susceptibility to microbial contamination that occurs during simulated handling of protective devices for the preparation of cytotoxic drug solutions.
METHODS AND RESULTS: Four devices, i.e. Chemoprotect spike, Clave connector, PhaSeal and Securmix were challenged with low and high inocula of micro-organisms. The cells, transferred to the connected vials during repeated manipulations of the devices were counted by means of solid-phase cytometry. Of the four devices, PhaSeal afforded the lowest transfer of micro-organisms. Secondly, the efficiency of procedures for the disinfection of an artificially contaminated rubber stopper was compared prior to connection of the vial to the PhaSeal device. Spraying or swabbing alone was inadequate, as opposed to a combination of spraying [0.5% or 2.0% (w/v) chlorhexidine in isopropanol] and swabbing [70% (v/v) isopropanol].
CONCLUSIONS: Although Phaseal afforded the lowest transfer of micro-organisms, adequate disinfection of the vial prior to connection remains required.
SIGNIFICANCE AND IMPACT OF THE STUDY: Unlike aspects of operator protection, which are well documented, the microbiological safety of protective devices for the preparation of cytotoxic drugs has not been addressed in the literature. This study estimates the susceptibility to microbial contamination during handling of four commonly used devices.
METHODS AND RESULTS: Four devices, i.e. Chemoprotect spike, Clave connector, PhaSeal and Securmix were challenged with low and high inocula of micro-organisms. The cells, transferred to the connected vials during repeated manipulations of the devices were counted by means of solid-phase cytometry. Of the four devices, PhaSeal afforded the lowest transfer of micro-organisms. Secondly, the efficiency of procedures for the disinfection of an artificially contaminated rubber stopper was compared prior to connection of the vial to the PhaSeal device. Spraying or swabbing alone was inadequate, as opposed to a combination of spraying [0.5% or 2.0% (w/v) chlorhexidine in isopropanol] and swabbing [70% (v/v) isopropanol].
CONCLUSIONS: Although Phaseal afforded the lowest transfer of micro-organisms, adequate disinfection of the vial prior to connection remains required.
SIGNIFICANCE AND IMPACT OF THE STUDY: Unlike aspects of operator protection, which are well documented, the microbiological safety of protective devices for the preparation of cytotoxic drugs has not been addressed in the literature. This study estimates the susceptibility to microbial contamination during handling of four commonly used devices.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app