Intra-abdominal vagal blocking (VBLOC therapy): clinical results with a new implantable medical device.
Surgery 2008 June
BACKGROUND: A new medical device uses high-frequency electrical algorithms to create intermittent vagal blocking (VBLOC therapy). The aim is to assess the effects of vagal blocking on excess weight loss (EWL), safety, dietary intake, and vagal function.
METHODS: An open-label, 3-center study was conducted in obese subjects (body mass index [BMI] 35-50 kg/m(2)). Electrodes were implanted laparoscopically on both vagi near the esophagogastric junction to provide electrical block. Patients were followed for 6 months for body weight, safety, electrocardiogram, dietary intake, satiation, satiety, and plasma pancreatic polypeptide (PP) response to sham feeding. To specifically assess device effects alone, no diet or exercise programs were instituted.
RESULTS: Thirty-one patients (mean BMI, 41.2 +/- 1.4 kg/m(2)) received the device. Mean EWL at 4 and 12 weeks and 6 months after implant was 7.5%, 11.6%, and 14.2%, respectively (all P < .001); 25% of patients lost >25% EWL at 6 months (maximum, 36.8%). There were no deaths or device-related serious adverse events (AEs). Calorie intake decreased by >30% at 4 and 12 weeks and 6 months (all P 25 pg/mL (P = .02). Three patients had serious AEs that required brief hospitalization, 1 each for lower respiratory tract, subcutaneous implant site seroma, and Clostridium difficile diarrhea.
CONCLUSIONS: Intermittent, intra-abdominal vagal blocking is associated with significant EWL and a desirable safety profile.
METHODS: An open-label, 3-center study was conducted in obese subjects (body mass index [BMI] 35-50 kg/m(2)). Electrodes were implanted laparoscopically on both vagi near the esophagogastric junction to provide electrical block. Patients were followed for 6 months for body weight, safety, electrocardiogram, dietary intake, satiation, satiety, and plasma pancreatic polypeptide (PP) response to sham feeding. To specifically assess device effects alone, no diet or exercise programs were instituted.
RESULTS: Thirty-one patients (mean BMI, 41.2 +/- 1.4 kg/m(2)) received the device. Mean EWL at 4 and 12 weeks and 6 months after implant was 7.5%, 11.6%, and 14.2%, respectively (all P < .001); 25% of patients lost >25% EWL at 6 months (maximum, 36.8%). There were no deaths or device-related serious adverse events (AEs). Calorie intake decreased by >30% at 4 and 12 weeks and 6 months (all P 25 pg/mL (P = .02). Three patients had serious AEs that required brief hospitalization, 1 each for lower respiratory tract, subcutaneous implant site seroma, and Clostridium difficile diarrhea.
CONCLUSIONS: Intermittent, intra-abdominal vagal blocking is associated with significant EWL and a desirable safety profile.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app