Ionic dialysance and the assessment of Kt/V: the influence of different estimates of V on method agreement.

BACKGROUND: Ionic dialysance was recently introduced as a means to assess Kt/V (K(ID)t/V). With this method, urea distribution volume (V) has to be estimated. The primary aim of the present study was to assess the agreement between equilibrated Kt/V assessed by urea kinetic modelling (eKt/V) with K(ID)t/V taking into account different estimates of V, and to assess the monthly variation in V. Secondly, the mechanisms behind the intra-treatment changes in ionic dialysance and inter-treatment variability of K(ID)t/V were assessed.

METHODS: Sixty-six patients were included. eKt/V was estimated using 30 min post-treatment sampling in the second generation Daugirdas equation. V was assessed by the formulae of Watson and Chertow (V(Watson); V(Chertow)), double-pool urea kinetic modelling (V(UKM)) and by ionic dialysance (V(IOD)) [Diascan; Hospal(R)].

RESULTS: The use of V(UKM) or V(IOD) instead of V(Watson) or V(Chertow) improved the relation between eKt/V and K(ID)t/V (both r = 0.93; P < 0.001 vs r = 0.84 and r = 0.81; P < 0.001). Mean values of eKt/V (1.19 +/- 0.21), K(ID)t/V(UKM) (1.19 +/- 0.30) and K(ID)t/V(IOD) (1.21 +/- 0.25) were comparable. Intra-class correlation coefficient of V(IOD) was 0.87 with a 1-month interval and <0.75 after 2 and 3 months. Intra-class correlation coefficient of V(DP) was 0.79 with a 1-month interval and <0.75 after 2 and 3 months. Inter-treatment variation in K(ID)t/V during six consecutive dialysis sessions was 6.1% +/- 0.6%. Changes in blood flow were the main determinant of variations in K(ID)t/V (P < 0.05). During treatment, ionic dialysance decreased by 12 +/- 13 ml/min (P < 0.001). The decline in blood volume was the major determinant of the intra-dialytic change in ionic dialysance (P < 0.05).

CONCLUSION: The use of V(IOD) and V(UKM) results in better agreement between eKt/V and K(ID)t/V compared with anthropometric formulae. K(ID)t/V was comparable with eKt/V and thus lower than expected for a single-pool method. V(IOD) and V(UKM), should be assessed at least monthly. K(ID)t/V varies widely between consecutive dialysis sessions, mainly due to differences in blood flow. During treatment, ionic dialysance decreases, which is related to the relative decline in blood volume.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app