Evaluation Studies
Journal Article
Add like
Add dislike
Add to saved papers

[Guidance of interventions in subintimal recanalization and fenestration of dissection membranes using a novel dual-lumen intravascular ultrasound catheter].

PURPOSE: To evaluate the feasibility and effectiveness of IVUS-guided puncture for gaining controlled target lumen reentry in subintimal recanalization of chronic iliac/femoral artery occlusions and in fenestration of aortic dissections.

MATERIALS AND METHODS: Between 5/2004 and 12/2005 12 consecutive patients (7 male, 5 female; mean age 64.6 +/- 12.0 years) with chronic critical limb ischemia and ischemic complications of aortic dissection were treated using the Pioneer catheter. This 6.2-F dual-lumen catheter combines a 20-MHz IVUS transducer with a pre-shaped extendable, hollow 24-gauge nitinol needle. This coaxial needle allows real-time IVUS-guided puncture of the target lumen and after successful reentry a 0.014" guidewire may be advanced through the needle into the target lumen. 7 patients were treated for aortic dissection and 5 patients (with failed previous attempts at subintimal recanalization) for chronic arterial occlusion. Patients with aortic dissection (5 type A dissections, 2 type B dissections) had developed renal ischemia (n = 2), renal and mesenteric ischemia (n = 2), or low extremity ischemia (n = 3). Patients with chronic arterial occlusions (2 common iliac artery occlusions, 3 superficial femoral artery occlusions) experienced ischemic rest pain (n = 4), and a non-healing foot ulcer (n = 1).

RESULTS: The technical success rate using the Pioneer catheter was 100%. The recanalization/fenestration time was 37 +/- 12 min. Procedure-related complications did not occur. In 10 cases a significant improvement of clinical symptoms was evident. One patient with aortic dissection and ischemic paraplegia required subsequent surgical intervention. One patient had persistent ischemic rest pain despite successful recanalization of a superficial femoral artery occlusion.

CONCLUSION: The Pioneer catheter is a reliable device which may be helpful for achieving target lumen reentry in subintimal recanalization of chronic occlusions and in fenestration of aortic dissections.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app