EVALUATION STUDIES
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

Comparison between different femoral fixation devices for ACL reconstruction with doubled hamstring tendon graft: a biomechanical analysis.

Arthroscopy 2006 June
PURPOSE: The aim of the present study was to compare the mechanical behavior of some devices used for femoral fixation of doubled hamstring tendon graft in anterior cruciate ligament (ACL) reconstruction when included in a graft fixation complex (GFC).

METHODS: An ACL reconstruction was performed on 90 porcine knees. The graft used was the doubled lateral extensor of toes (DLET). Nine different femoral fixation devices were tested and classified according to their fixation mechanism: compression (Bioscrew and RCI screw); expansion (Rigidfix); cortical suspension (Ligament Anchor, EndoButton-CL, and Swing Bridge); cancellous suspension (Linx-HT); and cortical-cancellous suspension (Transfix and Bio-Transfix). All GFC were subjected to a cyclic loading test, then to a load-to-failure test. Graft elongation after 1,000 load cycles, failure load, and stiffness were calculated for each device.

RESULTS: Regarding graft elongation, Bioscrew and RCI screws showed the highest mean values. All the other GFC showed no significant differences between them when delta elongation (elongation(1,000 cycles) - elongation(20 cycles)) was considered. For failure load, the highest mean values were observed for Bio-Transfix, Transfix, and Swing Bridge; a homogeneous subset with the lowest mean values was formed by Ligament Anchor, RCI screw, Bioscrew, and Linx-HT. For stiffness, the greatest values were observed for Bio-Transfix, Transfix, and Swing Bridge; all other groups showed no significant differences between them.

CONCLUSIONS: Cortical-cancellous suspension fixation seemed to offer the best and most predictable results in terms of elongation, fixation strength, and stiffness. For both compression and suspension, the weakest fixation was attained with cancellous fixation devices. Cortical suspension devices showed a greatly variable mechanical behavior, according to their design.

CLINICAL RELEVANCE: Transcondylar devices offer the best structural properties for femoral fixation of doubled hamstring tendon graft in ACL reconstruction.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app