Simulated clinical evaluation of four fluid warming devices*.

Anaesthesia 2006 June
The fluid warming capabilities of the Bair Hugger, Hotline, Standard Ranger and Fluido devices were evaluated in the laboratory with gravity flow via a 14G cannula (1 m head of fluid) and with the fluid bag pressurised to 300 mmHg. The resulting flows (70-450 ml.min(-1)) were recorded. At a room temperature of 22 degrees C, no device warmed the fluid to 37 degrees C. However, the Standard Ranger with gravity flow and the Fluido with both gravity and pressurised flow achieved 35 degrees C, whereas the Bair Hugger and Hotline with both gravity and pressurised flow, and the Standard Ranger with pressurised flow, achieved only 24-31 degrees C. However, from the way delivered temperatures changed with flow, we calculated that all four devices would achieve close to 37 degrees C at the flows specified by the manufacturers: 17, 83, 150 and 800 ml.min(-1) for the Bair Hugger, Hotline, Standard Ranger and Fluido, respectively.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app