Efficacy and safety results with the LifeSite hemodialysis access system versus the Tesio-Cath hemodialysis catheter at 12 months.

PURPOSE: To compare the performance and safety of a fully subcutaneous vascular access device, the LifeSite hemodialysis access system, versus a tunneled hemodialysis catheter, the Tesio-Cath, at 1 year after implantation.

MATERIALS AND METHODS: Sixty-eight patients who required hemodialysis received implantation of the LifeSite device or a Tesio-Cath device as a part of this multicenter study. Thirty-four patients were treated in each group. The endpoints observed included blood flow rates and associated venous pressures, overall and device-related adverse events, the need for thrombolytic infusions, device-related infections (DRIs) and associated hospitalizations, and technical device survival.

RESULTS: During the 12-month observation period, significantly higher venous pressures were required in patients with the Tesio-Cath to achieve blood flow rates comparable with those achieved with the LifeSite device. Patients in the LifeSite group experienced a significantly lower rate of non-device-related adverse events (P < .001), device-related adverse events (P < .016), need for thrombolytic infusions (P < .002), and DRIs (P < .013) compared with patients in the Tesio-Cath group. There was a trend toward a lower number of hospital days per month for DRIs in the LifeSite group, with the rate for the Tesio-Cath group being twice that in the LifeSite group. The use of the LifeSite device was also associated with a significantly higher probability of device survival for 12 months after censoring for planned removals (P < .031).

CONCLUSIONS: The results of the present study demonstrate superior device performance and technical device survival, reduced complications, and the need for fewer interventions with the LifeSite hemodialysis access system compared with a standard hemodialysis catheter during a 1-year time period after implantation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app