Myocardial tagging and strain analysis at 3 Tesla: comparison with 1.5 Tesla imaging.
Journal of Magnetic Resonance Imaging : JMRI 2006 April
PURPOSE: To determine whether imaging at 3 T could improve and prolong the tag contrast compared to images acquired at 1.5 T in normal volunteers, and whether such improvement would translate into the ability to perform strain measurements in diastole.
MATERIALS AND METHODS: Normal volunteers (N = 13) were scanned at 1.5 T (GE Signa CV/i) and 3.0 T (GE VH/i). An ECG-triggered, segmented k-space, spoiled-gradient-echo grid-tagged sequence was used during cine acquisition. Tag contrast was determined by the difference of the mean signal intensity (SI) of the tagline to the mean SI of the myocardium divided by the standard deviation (SD) of the noise (CNR(tag)). Matched short-axis (SA) slices were analyzed. Strain measurements were performed on images using a 2D strain analysis software program (harmonic phase (HARP)).
RESULTS: The average CNR(tag) over the cardiac cycle was superior at 3 T compared to 1.5 T for all slices (3 T: 23.4 +/- 12.1, 1.5 T: 9.8 +/- 8.4; P < 0.0001). This difference remained significant at cycle initiation, end-systole, and the end R-R interval (at cycle termination: 3 T = 14.0 +/- 11.0 vs. 1.5 T = 4.4 +/- 3.5; P < 0.01). Strain measures were obtainable only in early systole for 1.5 T images, but were robust throughout the entire R-R interval for 3 T images.
CONCLUSION: Imaging at 3 T had a significant benefit for myocardial tag persistence through the cardiac cycle. The improvement allowed strain analysis to be performed into diastole.
MATERIALS AND METHODS: Normal volunteers (N = 13) were scanned at 1.5 T (GE Signa CV/i) and 3.0 T (GE VH/i). An ECG-triggered, segmented k-space, spoiled-gradient-echo grid-tagged sequence was used during cine acquisition. Tag contrast was determined by the difference of the mean signal intensity (SI) of the tagline to the mean SI of the myocardium divided by the standard deviation (SD) of the noise (CNR(tag)). Matched short-axis (SA) slices were analyzed. Strain measurements were performed on images using a 2D strain analysis software program (harmonic phase (HARP)).
RESULTS: The average CNR(tag) over the cardiac cycle was superior at 3 T compared to 1.5 T for all slices (3 T: 23.4 +/- 12.1, 1.5 T: 9.8 +/- 8.4; P < 0.0001). This difference remained significant at cycle initiation, end-systole, and the end R-R interval (at cycle termination: 3 T = 14.0 +/- 11.0 vs. 1.5 T = 4.4 +/- 3.5; P < 0.01). Strain measures were obtainable only in early systole for 1.5 T images, but were robust throughout the entire R-R interval for 3 T images.
CONCLUSION: Imaging at 3 T had a significant benefit for myocardial tag persistence through the cardiac cycle. The improvement allowed strain analysis to be performed into diastole.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app