Evaluation Studies
Journal Article
Add like
Add dislike
Add to saved papers

Assessment of haemodialysis adequacy by ionic dialysance: intra-patient variability of delivered treatment.

INTRODUCTION: Adequate delivered dose of solute removal (as assessed by urea reduction and calculation of Kt/V) is an important determinant of clinical outcome in chronic haemodialysis (HD) patients. The requirement for multiple blood sampling and efforts taken to minimize the effects of rebound on post-treatment samples ensure Kt/V is measured only intermittently. On-line conductivity monitoring (using sodium flux as a surrogate for urea) allows the repeated non-invasive measurement of Kt/V on each HD treatment. We have studied the accuracy of this method of measuring Kt/V, and the variability of treatment dose delivered to individual patients.

METHODS: We prospectively studied 26 established chronic HD patients over 4 weeks (316 treatments). Patients were dialysed using Hospal Integra dialysis monitors, equipped with Diascan modules to measure Kt/V. Data were downloaded automatically to a central computer server. Urea reduction was measured (once a week) by a two-pool calculation using 30 min post-treatment sampling.

RESULTS: Treatment time, Q(B) and modality were fully delivered in all treatments analysed (97% of total). Kt/V measured by ionic dialysance (Kt/V(ID)) correlated highly with that derived from measurement of urea reduction (R(2)=0.92, P<0.0001). Kt/V(ID) underestimated urea-based Kt/V by a mean of only 1.5% (95% CI 0.18-2.9%). Kt/V(ID) varied greatly within individual patients with a mean CV of 0.13+/-0.10 (95% CI 0.05-0.3). If a Kt/V(ID) of 1.0 is considered 'adequate', 55% of the patients had variations that would have potentially altered their status as being adequately or inadequately dialysed, as the range of Kt/V readings cross that point during the study period.

CONCLUSION: In conclusion, Kt/V(ID) seems to be an accurate and readily obtained measure of adequacy. Substantial variation in Kt/V implies repeated measures (ideally for all treatments) are necessary to gain a true picture of the mean treatment dose being delivered to patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app