GUIDELINE
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

Implantable cardioverter defribrillator shocks: a troubleshooting guide.

Implantable cardioverter defibrillators deliver shocks in response to electrical signals that satisfy programmed criteria for detection of VT or VF. The first step in diagnosis of inappropriate shocks in patients with ICDs is to determine if the shock was delivered in response to a true tachyarrhythmia by inspecting data stored in the ICD. Shocks occur in the absence of tachyarrhythmias because nonarrhythmic physiologic or nonphysiologic signals are oversensed by the ICD and detected as arrhythmias. Diagnosis and causes of oversensing are reviewed. The second step in diagnosis is to determine if the tachyarrhythmia stored in the VT/VF episode log is VT/VF or SVT by analyzing stored electrograms. Frequent or repetitive shocks constitute an electrophysiologic emergency. The approach to this problem is reviewed.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app