Pacemaker troubleshooting: improved efficacy using a new diagnostic digital Holter recording system.

Asymptomatic dysfunction of cardiac pacing systems is not uncommonly detected using long-term ambulatory monitoring techniques. We report two patients with atrial and ventricular sensing abnormalities noted only on Holter monitoring. Multiple empirical attempts at pacemaker reprogramming based on deductive analysis of the pacing anomaly were unsuccessful. Through the use of a new digital recording system that allowed collection of surface electrocardiographic data, intracardiac electrograms, and Marker Channel data a correct diagnosis was made (pacing lead insulation failure). This new recording system has the advantage of significantly improving diagnostic sensitivity and resulting in cost savings.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app