BACKGROUND AND STUDY AIMS: Endoscopic perfusion manometry (PM) of the sphincter of Oddi (SO) requires expensive equipment which is relatively large and uncomfortable to handle in the endoscopic retrograde cholangiopancreatography (ERCP) setting. Furthermore, the volume load of the biliopancreatic system may contribute to the increased risk of pancreatitis after SO manometry.
PATIENTS AND METHODS: The newly developed small and lightweight microtransducer system consists of a portable data-logger with integrated online display which is connected to a 4-Fr manometry probe. The manometry catheter is inserted endoscopically into the biliopancreatic system via a 7-Fr Teflon sheath. SO motility can be observed online on the display but the data can also be stored for later analysis on a personal computer (PC). To validate the new method, 15 patients with suspected biliary SO dysfunction underwent both PM as well as microtransducer manometry (MTM) in randomized order. Thereafter, 50 consecutive patients with suspected biliary or pancreatic SO dysfunction were investigated solely by MTM.
RESULTS: PM was possible in 13 of 15 cases whereas MTM could be performed in all 15 patients. The basal SO pressure tended to be lower (approximately 5 mmHg) when measured with the MTM, compared with the PM method, but there was a significant and nearly linear correlation between the basal SO pressures obtained by both methods (r=0.98, P<0.001). SO dysfunction was diagnosed in the same five patients using both methods. Furthermore, the parameters of phasic SO motility were highly comparable when measured by MTM and PM. MTM was carried out successfully in 49 of 50 patients and only one MTM probe was used for all examinations, without malfunction. The endoscopist was able to diagnose SO dysfunction (by immediate observation of SO motility on the display) in 19 of 20 patients (when compared with the later PC analysis) and SO motility was judged correctly as normal in the remaining 29 cases. MTM was repeated in five patients with SOD 1-6 weeks after the first examination and the manometric findings were confirmed in all cases. Mild postmanometry pancreatitis was observed in only one of 49 patients (2%).
CONCLUSIONS: Endoscopic MTM is a reliable, safe, very easy to handle, and low-cost alternative for the clinical assessment of SO motility.
PATIENTS AND METHODS: The newly developed small and lightweight microtransducer system consists of a portable data-logger with integrated online display which is connected to a 4-Fr manometry probe. The manometry catheter is inserted endoscopically into the biliopancreatic system via a 7-Fr Teflon sheath. SO motility can be observed online on the display but the data can also be stored for later analysis on a personal computer (PC). To validate the new method, 15 patients with suspected biliary SO dysfunction underwent both PM as well as microtransducer manometry (MTM) in randomized order. Thereafter, 50 consecutive patients with suspected biliary or pancreatic SO dysfunction were investigated solely by MTM.
RESULTS: PM was possible in 13 of 15 cases whereas MTM could be performed in all 15 patients. The basal SO pressure tended to be lower (approximately 5 mmHg) when measured with the MTM, compared with the PM method, but there was a significant and nearly linear correlation between the basal SO pressures obtained by both methods (r=0.98, P<0.001). SO dysfunction was diagnosed in the same five patients using both methods. Furthermore, the parameters of phasic SO motility were highly comparable when measured by MTM and PM. MTM was carried out successfully in 49 of 50 patients and only one MTM probe was used for all examinations, without malfunction. The endoscopist was able to diagnose SO dysfunction (by immediate observation of SO motility on the display) in 19 of 20 patients (when compared with the later PC analysis) and SO motility was judged correctly as normal in the remaining 29 cases. MTM was repeated in five patients with SOD 1-6 weeks after the first examination and the manometric findings were confirmed in all cases. Mild postmanometry pancreatitis was observed in only one of 49 patients (2%).
CONCLUSIONS: Endoscopic MTM is a reliable, safe, very easy to handle, and low-cost alternative for the clinical assessment of SO motility.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app