Change in ventricular size and effect of ventricular catheter placement in pediatric patients with shunted hydrocephalus.

Neurosurgery 1999 December
OBJECTIVE: The multicenter, randomized pediatric cerebrospinal fluid shunt valve design trial found no difference in the rate of shunt failure between a standard valve, a siphon-reducing valve (Delta; Medtronic PS Medical, Goleta, CA), and a flow-limiting valve (Orbis Sigma; Cordis, Miami, FL); however, the valves were expected to have different effects on ultimate ventricular size. Also, the catheter position or local environment of the ventricular catheter tip might have affected shunt failure. Therefore, we performed a post hoc analysis to understand what factors, other than valve design, affected shunt failure and to identify strategies that might be developed to reduce shunt failure.

METHODS: Ventricular size was measured at as many as six different intervals, using a modified Evans' ratio (with incorporation of the frontal and occipital dimensions), in 344 patients. Ventricular catheter location was defined as being in the frontal horn, occipital horn, body of the lateral ventricle, third ventricle, embedded in brain, or unknown. The ventricular catheter tip was described as surrounded by cerebrospinal fluid, touching brain, or surrounded by brain parenchyma within the ventricle (slit ventricle). Repeated measures analysis of variance for unbalanced data was used to analyze ventricular size. A Cox model (with incorporation of time-dependent covariates) was used to evaluate the contribution of age, etiology, shunt design, ventricular size, ventricular catheter location, and environment among the cases.

RESULTS: Ventricular volume decreased in an exponential fashion, forming a plateau at 14 months, and was similar for the three valves (P = 0.4). Frontal and occipital ventricular catheter tip locations were associated with a reduced risk of shunt failure (hazard ratios, 0.60 [P = 0.02] and 0.45 [P = 0.001], respectively). Ventricular catheter tips surrounded by cerebrospinal fluid or touching the brain were associated with a reduced risk of failure (hazard ratios, 0.21 and 0.33, respectively; P = 0.0001). Patients with myelomeningocele or large ventricles had increased risk of malfunction (hazard ratios, 1.78 [P = 0.006] and 2.33 [P = 0.03], respectively).

CONCLUSION: Decline of ventricular size over time is not affected by these different shunt valve designs. This suggests that the mechanical models of hydrocephalus on which the designs were based are inadequate. Ventricular catheter tip location and ventricular catheter environment are important. Techniques to accurately place ventricular catheters and new valve designs that effectively control ventricular size might reduce shunt malfunction.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app